Table 9-6 List of self-test errors (continued)
645
646
647
Calibration errors
The following errors indicate failures that may occur during a calibration.
Table 9-7 List of calibration errors
701
703
705
707
708
709
720
721
722
723
726
742
743
744
745
U3606A User's and Service Guide
Error code
Error message
Constant Current 0.2V/10kohm or amplified 11 error
Constant Current 0.8V/100kohm or amplified 11 error
Constant Current 0.8V/1.1Mohm or amplified 11 error
Error code
Error message
Cal security pads short
Invalid secure code
Cal aborted
Cal signal measurement out of range
Cal signal frequency out of range
EEPROM write failure
Cal DCV offset out of range
Cal DCI offset out of range
Cal RES offset out of range
Cal CAP offset out of range
Cal RES open out of range
Cal checksum failed, DCV corrections
Cal checksum failed, DCI corrections
Cal checksum failed, RES corrections
Cal checksum failed, ACV corrections
List of Error Messages
Error Messages
9
263