Rtd Measurements; Thermistor Measurements - Agilent Technologies 34420A User Manual

Nano volt/micro ohm meter
Table of Contents

Advertisement

Chapter 3 Features and Functions
Temperature Measurement Configuration

RTD Measurements

The meter supports RTDs with
= 0.00391. The nominal value of the RTD (R
of 4.9
to 2.1 k . RTD Connections are shown on page 21.
The meter makes a 4-wire measurement for RTD's. All four input
leads must be connected.
Front Panel operation: Set RTD as the probe type under the
1: PROBE TYPE command in the TEMPerature menu.
Select the RTD type under the 3: RTD TYPE command in the
TEMPerature menu.
Enter the 0° C resistance of the RTD (R
in the TEMPerature menu.
Remote operation: Use the following commands for temperature
measurements with RTD's:
CONFigure:TEMPerature
MEASure:TEMPerature?
SENSe:FUNCtion "TEMPerature"
SENSe:TEMPerature:TRANsducer:TYPE
SENSe:TEMPerature:TRANsducer:FRTD:TYPE
SENSe:TEMPerature:TRANsducer:FRTD:RESistance

Thermistor Measurements

Only the 5 k thermistor type is supported. Thermistor connections
are shown on page 21.
The thermistor must be connected to the Channel 1 inputs.
Front Panel operation: Set THERM as the transducer type under
the 1: PROBE TYPE command in the TEMPerature menu.
Remote operation: use the following commands for temperature
measurements with a thermistor:
CONFigure:TEMPerature
MEASure:TEMPerature?
SENSe:FUNCtion "TEMPerature"
SENSe:TEMPerature:TRANsducer:TYPE
72
= 0.00385 (DIN/IEC 751) or
) must be in the range
0
) in the 4: RTD R
0
FRTD,{85|91}[,1,< resolution >]
FRTD, {85|91}[,1,< resolution >]
FRTD
THER, DEF,[,1,< resolution >]
THER, DEF,[,1,< resolution >]
THER
command
0
{85|91}
< value >

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents