Special Resistance Measurement Commands - Agilent Technologies 34420A User Manual

Nano volt/micro ohm meter
Table of Contents

Advertisement

Chapter 4 Remote Interface Reference

Special Resistance Measurement Commands

Special Resistance Measurement Commands
The following commands provide offset compensation and low power or
voltage limited resistance measurement capability.
[SENSe:]
FRESistance | RESistance
:OCOMpensated {OFF | ON}
:OCOMpensated?
This command enables or disables the offset compensated resistance
measurements (see page 69). Once enabled, offset compensation will
apply to both 4-wire and 2-wire resistance measurements.
[SENSe:]
FRESistance
:POWer:LIMit[:STATe] {OFF | ON}
:POWer:LIMit[:STATe]?
This command enables or disable the 4-wire low power resistance mode
(see page 69).
[SENSe:]
FRESistance
:VOLTage:LIMit[:STATe] {OFF | ON}
:VOLTage:LIMit[:STATe]?
:VOLTage:LIMit:VALue {< value > | MIN | MAX}
:VOLTage:LIMit:VALue? {MIN | MAX}
This command enables or disables the 4-wire voltage limited resistance
mode (see page 70). The voltage limit <value> parameter will be rounded
to one of three values:
20 mV,
100 mV, or
500 mV
MIN sets 20 mV and MAX sets 500 mV.