Chapter 4 Remote Interface Reference

Input Message Terminators

Input Message Terminators
Program messages sent to a
<newline> character. The
interpreted as a <newline> character and may also be used to terminate
a message in place of the <newline> character. A <carriage return>
followed by a <newline> is also accepted. Many programming languages
allow you to specify a message terminator character or
automatically sent with each bus transaction. Message termination will
always set the current path back to the root-level.

Output Data Formats

Output data will be in one of formats shown in the table below.
Type of Output Data
Non-reading queries
Single reading (IEEE-488)
Multiple readings (IEEE-488)
Single reading (RS-232)
Multiple readings (RS-232)
186
instrument must terminate with a
SCPI
(end or identify) signal is
IEEE-488 EOI
Output Data Format
<80 ASCII character string
SD.DDDDDDDDESDD<nl>
SD.DDDDDDDDESDD,...,... <nl>
SD.DDDDDDDDESDD<cr><nl>
SD.DDDDDDDDESDD,...,...<cr><nl>
S Negative sign or positive sign
D Numeric digits
E Exponent
<nl> newline character
<cr> carriage return character
state to be
EOI