Calibration Errors

The following errors indicate failures that may occur during a calibration.
In error 732, "xxxx" is the frequency at which the worst AC flatness deviation was
N O TE
calculated.
34410A/11A/L4411A Service Guide
Error #
Meaning
701
Calibration error; security defeated by hardware jumper
702
Calibration error; calibration memory is secured
703
Calibration error; secure code provided was invalid
704
Calibration error: secure code too long
705
Calibration error; calibration aborted
706
Calibration error: value out of range
707
Calibration error: signal measurement out of range
708
Calibration error: signal frequency out of range
709
Calibration error: no cal for this function or range
710
Calibration error: full scale correction out of range
711
Calibration error: ADC calibration failed
720
Calibration error: DCV offset out of range
721
Calibration error: DCI offset out of range
722
Calibration error: RES offset out of range
726
Calibration error: ACV offset out of range
727
Calibration error: ACI offset out of range
730
Calibration error: precharge DAC convergence failed
731
Calibration error: A/D turnover correction out of range
732
Calibration error: AC flatness calibration failed; amplitude at xxxx Hz
733
Calibration error: AC low frequency convergence failed
734
Calibration error: AC low frequency correction out of range
747
Calibration variable does not exist
748
Cal: mainframe cal memory write failure