Measurement Characteristics - Agilent Technologies 34420A User Manual

Nano volt/micro ohm meter
Table of Contents

Advertisement

DC Voltage Noise Specifications *
Observation Period
2-Minute
RMS
Range
Noise
1 mV
1.3 nV
RMS
10 mV
1.5 nV
RMS
100 mV
10 nV
RMS
1 V
100 nV
RMS
10 V
450 nV
RMS
100 V
11 V
RMS
* After a 2-hour warm-up, ± 1°C, 6.5 digits (10 PLC) with Analog
Filter Off and Digital Filter Medium (50 reading average). 2-Minute
RMS and 24-hour noise typical. For measurements using 0.02 or
0.2 NPLC, add 800 nV RMS noise.
DC Voltage Noise vs Source Resistance *
Source
Noise
Resistance
0
1.3 nV
RMS
100
1.7 nV
RMS
1k
4 nV
RMS
10k
13 nV
RMS
100k
41 nV
RMS
1M
90 nV
RMS
* Typical noise behavior for Ch 1 or Ch 2, after 2 hour warm-up,
6.5 digits (10 PLC), 2 minute observation period on 1 mV range.
For peak-to-peak noise, multiply RMS noise by 6.
Chapter 8 Specifications
34420A Nano Volt/Micro Ohm Meter
2-Minute
24-Hour
Peak-Peak
Peak-Peak
Noise
Noise
8 nVpp
12 nVpp
10 nVpp
14 nVpp
65 nVpp
80 nVpp
650 nVpp
800 nVpp
3 Vpp
3.7 Vpp
75 Vpp
90 Vpp
Analog Filter Digital Filter
Off
Med
Off
Med
Off
Med
Off
Med
On
Med
On
Slow

MEASUREMENT CHARACTERISTICS

DC Voltage
Measurement Method
Continuously Integrating
Multi-slope III A-D Converter
A-D Linearity
0.00008% of reading + 0.00005% of range
Input Resistance
100V (Ch1 only)
10 M ± 1%
1mV through 10V
>10 G , in parallel with <3.6 nF
Input Bias Current
<50 pA at 25°C
Injected Current*
<50 nA pp at 50 or 60 Hz
Input Protection
150 V peak any input terminal to
Channel 1 LO, continuous
Channel Isolation
Isolation between input channels
>10
Earth Isolation
350 V peak any input terminal to
earth. Impedence from any input
terminal to earth is
>10 G and < 400 pF
Maximum Voltage
Channel 1 LO to Channel 2 LO,
150V peak
Resistance
Measurement Method
Selectable 4-wire or 2-wire ohms.
Current Source referenced to
Channel 1 LO input
Offset Compensation
Used on all ranges except
100 k and 1 M .
Can be turned off if desired
Protection
150V peak
Open Circuit Voltage
For Resistance and Low Power
Resistance <14 V. 20 mV, 100 mV,
500 mV selectable clamp
* Injected current is calibrated at the factory. Different line power
configurations will require recalibration.