Chapter 4 Remote Interface Reference

Measurement Configuration Commands

Measurement Configuration Commands
See also "Measurement Configuration," starting on page 51 in chapter 3.
FUNCtion "<
function
>"
Select a measurement function. The function must be enclosed in quotes
in the command string (FUNC "VOLT:DC"). Specify one of the following
strings.
VOLTage:DC
VOLTage:DC:RATio
VOLTage:AC
CURRent:DC
CURRent:AC
2-wire ohms
RESistance
(
FUNCtion?
Query the measurement function and return a quoted string.
<
function
>:RANGe {<
range
Select the range for the selected function. For frequency and period
measurements, ranging applies to the signal's input voltage, not its
frequency (use FREQuency:VOLTage or PERiod:VOLTage).
the lowest range for the selected function.
range. [Stored in volatile memory]
function
<
>:RANGe? [MINimum|MAXimum]
Query the range for the selected function.
function
<
>:RANGe:AUTO {OFF|ON}
Disable or enable autoranging for the selected function. For frequency
and period, use FREQuency:VOLTage or PERiod:VOLTage.
Autorange thresholds: Down range at <10% of range; Up range
at >120% of range. [Stored in volatile memory]
<
function
>:RANGe:AUTO?
Query the autorange setting. Returns "0" (
4-wire ohms
FRESistance (
FREQuency
PERiod
CONTinuity
DIODe
)
>|MINimum|MAXimum}
MAX
OFF
)
selects
MIN
selects the highest
) or "1" (
).
ON
4
121