To Make a 2-Wire Resistance Measurement

Press
• Ranges: 100 W, 1 kW, 10 kW, 100 kW, 1 MW, 10 MW, 100 MW, 1 GW
• Configurable parameters: INTEGRATION, RANGE, OFFSET COMP, AUTO ZERO, NULL,
To null–out the test lead resistance:
1 Connect one end of the test leads at the meter, and short the probe
2 Press null.
3 Connect the probe ends to the test circuit, and measure the corrected

To Make a 4-wire Resistance Measurement

Press
• Ranges: 100 W, 1 kW, 10 kW, 100 kW, 1 MW, 10 MW, 100 MW, 1 GW
• Configurable parameters: INTEGRATION, RANGE, OFFSET COMP, NULL, and NULL
All 4-wire resistance measurements are made with auto-zero on.
44
to select the 2-wire resistance function.
and NULL VALUE
Connect test leads as shown:
ends together.
resistance value.
VALUE
Connect test leads as shown:
to select the 4-wire resistance function.
Resistance
Resistance