4.5.3 Auto Cal Screen

When proper automatic valving is connected, the analyzer can
cycle itself through a sequence of steps that automatically zero and span
the instrument. The Auto-Cal screens that follow can be used to define
an automatic calibration sequence and/or start an Auto-Cal event.
The Auto Cal function is available from the first System screen.
To enter the Auto Cal function press System and then Auto Cal. A
series of screens will appear beginning with the following screen:
Auto Cal screen has the following options,
1. Auto Cal Zero ON or OFF
2. Auto Cal Span ON or OFF
The Zero/Span schedule can be set to xxx days and xx hours before
the next auto calibration cycle will automatically occur. The Zero and
Span rows have Days and Hours Edit boxes. Clicking on the edit box
will bring up individual number pads for entering or editing the Days
and Hours setting.
The Zero Days / Span Days range from 1 to 999.