Agilent Technologies 1260 Infinity VL User Manual page 40

Quaternary pump
Table of Contents

Advertisement

3
Installing the Pump
Flow Connections of the Pump
1 Remove the front cover by pressing the snap fasteners on both sides.
2 Place the solvent cabinet on top of the quaternary pump.
3 Put the bottle-head assemblies into empty solvent reservoirs and place the
4 Connect the inlet tubes from the bottle-head assemblies to the inlet
5 Connect the solvent tubes from the MCGV inlet to the outlets of the vacuum
6 Using a piece of sanding paper connect the waste tubing to the purge valve
7 If the pump is not part of an Agilent 1260 Infinity system stack or placed on
8 Connect the pump outlet capillary (pump to injection device) to the outlet
9 Fill solvent reservoirs with your mobile phase.
40
Figure 7
Removing the Front Corver
bottle in the solvent cabinet.
connectors A to D at the right hand side of the vacuum degasser, see
Figure 8
on page 41. Fix the tubes in the tube clips of the pump.
degasser.
and place it into your waste system.
the bottom of a stack, connect the waste tube to the waste outlet of the
pump leak handling system.
of the purge valve.
1260 Infinity QuatPump VL User Manual