Temperature Limit Exceeded

Error ID: 2517
The temperature of one of the motor-drive circuits is too high.
The processor continually monitors the temperature of the drive circuits on
the main board. If excessive current is being drawn for long periods, the
temperature of the circuits increases. If the temperature exceeds the upper
limit, the error message is generated.
Probable cause
1
2
3
4
1260 Infinity QuatPump VL User Manual
High friction (partial mechanical blockage) in
the pump drive assembly.
Partial blockage of the flowpath in front of
the damper.
Defective pump drive assembly.
Defective main board.
Error Information
Module Error Messages
Suggested actions
Remove the pump-head assembly. Ensure there
is no mechanical blockage of the pump-head
assembly or pump drive assembly.
Ensure the outlet valve is not blocked.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.