Running The Test; Evaluating The Results - Agilent Technologies 1260 Infinity VL User Manual

Quaternary pump
Table of Contents

Advertisement

8
Test Functions and Calibration
Leak Rate Test

Running the Test

Parts required
p/n
01080-83202
Running the test from the Agilent Lab Advisor
1 Select the Leak Rate Test from the Test Selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by slowly opening the purge valve when the test has
N O T E
finished.

Evaluating the Results

Results of the leak rate test are the leak rates measured for pistons 1 and 2 as
described for the test principle. If any of the leak rates exceeds 3 µL/min, the
test will fail.
104
Description
Blank nut
1260 Infinity QuatPump VL User Manual