Forced Cold Start
A forced cold start can be used to bring the module into a defined mode with
default parameter settings.
Loss of data
C A U T I O N
Forced cold start erases all methods and data stored in the non-volatile memory.
Exceptions are calibration settings, diagnosis and repair log books which will not be
erased.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
Table 19
Forced Cold Start Settings (without on-board LAN)
Mode Select
TEST/BOOT
1260 Infinity QuatPump VL User Manual
Setting the 8-bit Configuration Switch (without On-board) LAN
Save your methods and data before executing a forced cold start.
SW1
SW2
SW3
1
1
SW4
SW5
0
0
1
Hardware Information
SW6
SW7
SW8
0
0