Error Information
Module Error Messages

Wait Timeout

Error ID: 2053
When running certain tests in the diagnostics mode or other special
applications, the pump must wait for the pistons to reach a specific position,
or must wait for a certain pressure or flow to be reached. Each action or state
must be completed within the timeout period, otherwise the error message is
generated.
Possible Reasons for a Wait Timeout:
· Pressure not reached.
· Pump channel A did not reach the delivery phase.
· Pump channel B did not reach the delivery phase.
· Pump channel A did not reach the take-in phase.
· Pump channel B did not reach the take-in phase.
· Solvent volume not delivered within the specified time.
Probable cause
1
2
3
4
92
Purge valve open.
Leak at fittings, purge valve, active inlet
valve, outlet valve or piston seals.
Flow changed after starting test.
Defective pump drive assembly.
Suggested actions
Ensure that purge valve is closed.
Ensure pump components are seated correctly.
If there are still signs of a leak, exchange the
appropriate seal (purge valve, active inlet valve,
outlet valve, piston seal).
Ensure correct operating condition for the
special application in use.
Please contact your Agilent service
representative.