Remote Cables - Agilent Technologies 1260 Infinity VL User Manual

Quaternary pump
Table of Contents

Advertisement

11
Identifying Cables

Remote Cables

Remote Cables

One end of these cables provides a Agilent Technologies APG (Analytical
Products Group) remote connector to be connected to Agilent modules. The
other end depends on the instrument to be connected to.
Agilent Module to 3396A Integrators
p/n 03394-60600
Agilent Module to 3396 Series II / 3395A Integrators
Use the cable Agilent module to 3396A Series I integrators (p/n 03394-60600)
and cut pin #5 on the integrator side. Otherwise the integrator prints START;
not ready.
160
Pin 3396A
Pin Agilent
module
9
1 - White
NC
2 - Brown
3
3 - Gray
NC
4 - Blue
NC
5 - Pink
NC
6 - Yellow
5,14
7 - Red
1
8 - Green
NC
9 - Black
13, 15
1260 Infinity QuatPump VL User Manual
Signal Name
Active
(TTL)
Digital ground
Prepare run
Low
Start
Low
Shut down
Low
Not connected
Power on
High
Ready
High
Stop
Low
Start request
Low
Not connected

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1311c

Table of Contents