9
Maintenance
Exchanging the Purge Valve Frit
6
Place a new frit into the purge valve with the orientation
of the frit as shown below (slit in frit points to the front).
Reinstall the seal cap including the gold seal.
7
Insert the purge valve into the pump head and locate the
pump outlet capillary and the waste tube.
Outlet capillary
Purge valve
Waste tube
118
N O T E
Before reinstallation always check the gold seal in the
seal cap. A deformed seal cap should be exchanged.
8
Tighten the purge valve and reconnect outlet capillary and
waste tubing.
1260 Infinity QuatPump VL User Manual