GE Definium 5000 Operator's Manual page 186

X-ray system
Table of Contents

Advertisement

Recommendation: When building a custom look, start with the factory look that is closest to the desired
result. Write down the values and apply them to a new custom process. Then change the applicable
variables.
1. From the Worklist screen, press [UTILITIES].
2. Select Preferences > Image Processing.
3. Press Image Processing [EDIT].
The Image Processing Preference Editor (Figure 10-24) appears.
Figure 10-24 Image Processing Preference Editor
4. Select the Anatomy (for example: Chest, Abdomen, Lumbar Spine, etc.).
5. Select the View (for example: antero-posterior or lateral).
6. Select the Image Type (some systems may only have Standard image type available).
7. Select the Patient Size. Available options are: Small Adult, Medium Adult, Large Adult, Small Pediatric,
Medium Pediatric, or Large Pediatric.
8. Select the Look. Choose the first "Undefined" Custom look.
NOTE: If a Factory Look is selected, parameters are read-only. Only Custom processing looks can be
changed.
9. Enter new name in Look Description to rename the look:
The system will not accept a new look if the name remains "Undefined."
10. Check the Make Default box to set this look at the default for the selected Anatomy.
11. Adjust parameters as described in Table 10-10. Parameters may be adjusted in any order.
Definium 5000 X-Ray System
Operator Manual
5197809-1EN
Rev. 4 (13 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter 10: Set Preferences
10-34