Factors
Factors
Maximum Power kW
Maximum mA
Maximum kVp
Power Output
(@0.1 s)

Duty Cycle

18 exposures per hour at maximum mAs, using the X-ray tube: Toshiba E7254FX.

Generator Accuracy

Tube Voltage Range
Tube Current Range
Accuracy of Exposure Amount
(mAs)
Time of Exposure

X-Ray Source Assembly Filtration

The x-ray source assembly is made up of the x-ray tube and the collimator. Together, they provide
permanent, non-removable filtration of 3.6mm aluminum equivalent @ 80 kVp.
Definium 5000 X-Ray System
Operator Manual
65 kW
550 mA
150 kVp
650 mA @ 40-98 kVp
500 mA @ 99-128 kVp
400 mA @ 129-150 kVp
X-Ray voltage range: 40kVp - 150kVp in 1 kVp steps
Allowable deviation: +/- (3% + 1kV)
X-Ray tube current range: 10 - 650 mA (65kW)
Allowable deviation: +/- (4% + 1mA)
0.1 mAs - 520 mAs
Manual mode maximum mAs limited to 520 mAs
AEC mode maximum mAs limited to 512 mAs
1ms - 2000ms
Allowable deviation: +/- (2% + 0.1ms)
5197809-1EN
Rev. 4 (13 February 2008)
© 2008 General Electric Company. All rights reserved.
Appendix C: Technical Specif ications
Generator SHF-635