3. Press Default Factory Looks [EDIT].
4. Use the drop-down lists to change the default for the desired anatomical views.
5. When finished, click [MAP] to save the changes.
A message appears: "Default look settings for selected Representative Anatomical Views will be
mapped to related anatomical views."
6. Click [CONFIRM] to apply the looks settings. [CONFIRM] will close the message and return you to the
Image Processing screen.
[CANCEL] closes the message and returns you to the Default Factory Looks screen.
Figure 10-23 Default Factory Looks

Build Custom Looks

Looks are the way an image is processed to be viewed by a radiologist for interpretation.
Looks are pre-defined combinations of brightness, contrast, edge enhancement, and tissue equalization
(TE). Refer to
Tissue Equalization Overview
Factory Look descriptions:
Factory Look 1 – similar to analog film, low edge, no TE
Factory Look 2 – low CR look, low edge, moderate tissue contrast
Factory Look 3 – moderate CR look, low edge, TE
Factory Look 4 – highly digital look, high edge, high TE
Follow this process to create an new custom look.
Definium 5000 X-Ray System
Operator Manual
(p. 10-37) for more information.