Table 3-3 Hand switch button positions
Position
OFF
PREP
EXPOSE

System Movements

WARNING:
Monitor system movements to avoid any impact of the system with the floor, ceiling, or
other objects in the room. Impact may cause serious damage to the equipment.
WARNING:
Monitor the patient position (hands, feet, fingers, etc.) to avoid injury to patient caused by
unit movements. Patient must be kept away from mobile components of the unit.
Intravenous tubing, catheters, and other patient connected lines should be routed away
from moving equipment.
Definium 5000 X-Ray System
Operator Manual
The OFF position is when no pressure is applied to the Prep/Expose
button on top of the hand switch.
PREP is the next position on the hand switch. When it is partially
pressed, it brings the rotor up to speed and heats the filament. The
button returns to the OFF position if released.
The EXPOSE position is when the button on the hand switch is fully
depressed. This produces X-rays that are recorded. Release the Prep/
Expose button after the exposure is completed.
5197809-1EN
Rev. 4 (13 February 2008)
© 2008 General Electric Company. All rights reserved.
Chapter 3: System Overview
Description
3-7