Honeywell SmartLine Multivariable SMV800 Series User Manual page 290

Hart/de option
Hide thumbs Also See for SmartLine Multivariable SMV800 Series:
Table of Contents

Advertisement

5
6
7
8
9
10
Page 278
Enter the relevant flow process data from the Averaging Pitot Tube
Sizing Data Sheet into the appropriate entry fields on the Process Data
page as follows:
Normal Flowrate
Normal DP = 13.21 inches H
Design Density
You can change the engineering units by clicking on the text box with
right mouse button.
to proceed to the Flowing Variables page.
Next
Click the following options for failsafe indication on the Flowing
Variables page (so that there is an "
This will ensure that the PV4 flow output will go to failsafe if either
the static pressure or temperature sensors fail.
Set Damping = 1.0 seconds.
Click Next to proceed to the Solutions page.
The calculated Kuser value appears on the Solutions page of the
Kuser Model along with a list of items (with values) that you have
configured from previous pages. Review the Wizard values to make
sure they are correct.
Click Finish to complete the Kuser calculation procedure.
Connect SCT to SMV and establish communications. (Refer to the
SCT manual #34-CT-10-08 for procedure, if necessary.)
Perform Download of the database configuration file to the SMV.
Use the procedure in section 0, to verify the Kuser and flow
calculation for this application.
You can simulate inputs for PV1, PV2, and PV3 to verify PV4 output.
SMV800 Series HART/DE Option User's Manual
= 45,000 lb/hr
O @ 39.2 °F
2
3
= 0.49659 lbs/ft
" in each check box):
a
Revision 6.0