8.4.2.3

Configuration: Special functions: ELAN/PROFIBUS/external interference

ELAN parameters
ELAN ext. interfer.
PROFIBUS parameters
Const. interference
Configuration: Special functions: ELAN/PROFIBUS/external interference: ELAN parameters
Channel
Name :ULTTWENTYTHREE
Protocol
Sync.zero cal.
Note
For further details on operation of the analyzer in an ELAN network, please refer to the ELAN
interface description (C79000-B5274-C176 German/English).
Configuration: Special functions: ELAN/PROFIBUS/external interference: ELAN external
interference
Interference CO2
Interferences : OFF
Interference 1
Interference 2
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012
You can use this dialog to configure the analyzer for use in an
ELAN or PROFIBUS network.
The adjacent menu display appears when you select the function
from the higher-level menu.
8423
You can use this dialog to set the parameters for an ELAN net‐
:
1
work. These are:
:OFF
• Channel
:OFF
You must set the channel addresses for the analyzers present
84231
in an ELAN network. Addresses from 1 to 12 can be set, where
each address may only be used once.
• Name
You can set an analyzer name here. During communication
with ELAN, it can be used for the plain text identification of the
analyzers. An analyzer name may have up to 10 alphanumeric
characters.
• Protocol (ON/OFF)
The automatic transmission of measured values can be switch‐
ed on/off. With 'ON', the analyzer sends a measured value
frame cyclically every 500 ms.
Note
To avoid considerably hindering communication within an
ELAN network, this function should only be set to 'ON' when
required (e.g. with correction of cross-interference).
• Sync. zero cal. (ON/OFF)
This function is not yet available. Therefore only 'OFF' is the
currently valid value.
This function can be used for a correction calculation by measur‐
ing the influence of an interfering gas by means of another ana‐
lyzer connected in the ELAN network.