Description

3.2 Design

3.2
Design
Figure 3-2
The ULTRAMAT 23 is also available as a bench-top unit. This version differs from the rack unit
shown here as follows:
• Closed housing without mounting frame.
• 2 handles on the sides.
• The bench-top unit is not available with Ex approval.
• Condensation trap with safety filter.
Enclosure
• Bench-top unit or
• 19" rack unit with 4 HU for installation in
– Hinged frame
– Cabinets
• Flow indicator for sample gas on front plate (not with piped gas paths)
• Integrated sample gas pump with bench-top unit, available as option for rack unit (not with
piped gas paths and not in combination with the UV module).
• Gas connections for sample gas inlet and outlet as well as zero gas possible with pipe
diameter 6 mm or ¼"
22
Available as bench-top or rack unit
80-digit display panel (4 lines with 20 characters each)
Dust-tight and washable membrane keyboard
Flowmeter for monitoring the sample gas flow
Gas and electrical connections at the rear
Warning label
O
sensor (option) can be removed from the front after the cover has been taken off
2
Machine-readable product designation
Control and function keys for operation
Design of ULTRAMAT 23 as 19" rack unit