Ir Analyzer Unit - Siemens ULTRAMAT 23 Series Equipment Manual

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Description
3.4 Technical specifications
Operation of analyzers in hazardous areas
For safe operation of the analyzer in hazardous areas, it is essential to observe the information
and conditions referred to in the "ATEX compact operating instructions for ULTRAMAT 23".
3.4.2

IR analyzer unit

To ensure compliance with the technical specifications, a cycle time of ≤ 24 hours must be
activated for the AUTOCAL. The technical specifications are based on a sample gas pressure of
1013 ±5 hPa absolute, a sample gas flow of 1.2 ±0.2 l/min and an ambient temperature of
25 ±2 °C.
General information
Measured components
Measuring ranges
Display resolution
Chopper compartment purging
Inlet pressure
Purging gas consumption
Time response
Warm-up period
Response time (T
time)
90
Damping (electronic time constant)
Measuring behavior at analog output
Output signal noise
Output signal resolution
Linearity error
Repeatability
34
WARNING
Maximum of 3
See ordering data
Depends on the selected measuring range
Approx. 3000 hPa (43.5 psi) absolute
Approx. 100 ml/min
Approx. 30 minutes at room temperature. The measured values are invalid
for the first 30 minutes following switch-on. The maximum accuracy is
achieved after approx. 2 hours
Dependent on length of analyzer chamber, sample gas feed line
and parameterizable attenuation
0 ... 99.9 s, adjustable
<1% of current measuring range (or smallest possible measuring range
according to nameplate)
< 0.1% of the output signal span
In the largest measuring range: < ±1% of the full scale value
In the smallest measuring range: < 2% of the full scale value
≤ 1% of the current measuring range
Equipment Manual, 04/2022, A5E37100388-012
ULTRAMAT 23