Siemens ULTRAMAT 23 Series Equipment Manual page 141

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Function
CAL/MEAS
Function control
External solenoid valves
Zero gas
Calibration gas (sample
gas inlet)
Sync.
The pin assignments of the relays in the de-energized state are described in section Pin
assignments (Page 53).
The functions which can be assigned to the relays have the following meanings:
Relays SO2
Limit 1: Relay 1
Limit 2: Relay 2
NO
CO
SO2
O2
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012
Relay de-energized
Measuring; warm-up
phase (5 min); purging
function of the H
S sensor;
2
protection function of the
H
S sensor
2
Function control is present
Zero gas flows;
AUTOCAL gas flows (not
during AUTOCAL sample
gas pre-purging phase);
purging function of H
S
2
sensor; protection func‐
tion of H
sensor
2
limits
84121
MR relay
3
MR relay
-
MR relay
4
MR relay
-
84122
Relay energized
AUTOCAL; all calibrations
Calibration gas/scan gas flows
(valid for IR components and
all sensors)
Synchronization signal is out‐
put
Limits
Upper and lower limits can be defined as events for energizing
relays. Select the desired relay(s) in the second and third lines of
this menu.
This function is specific to the component.
Measuring range
A relay for measuring range switchover can be assigned to each
measured component. This guarantees reliable assignment of
the analog output signal to the currently active range, especially
with autoranging (see section Parameter: Measuring ranges
(Page 127)).
8.4 Configuration
Signaling
AUTOCAL; all calibrations
Remote operation
During warm-up phase (approx.
30 min, longer with installed UV
module), AUTOCAL, uncoded
Ext. solenoid valve open
Ext. solenoid valve open
AUTOCAL only "Zero gas flow" and
adjustment (not during sample gas
pre-purging phase);
calibration gas for the AUTOCAL
flows (not during the sample gas pre-
purging phase)
Functions
141