Siemens ULTRAMAT 23 Series Equipment Manual page 142

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Functions
8.4 Configuration
Fault
Maintenace req.:R5
CAL/MEAS
FunctionControl:R-
External valves
zero gas : relay -
span gas : relay 7
Sync
142
Status messages
:R-
You can use this menu to apply the signaling of various operat‐
:R6
ing states of the analyzer as an event for relay control
(R in the display means Relay).
84123
The following signals are possible:
• Fault:
Occurrence of a fault and output of a fault message
• Maintenance request:
Occurrence of maintenance request
(assigned to relay 5 in Fig.)
• CAL/MEAS:
Switching over from measuring mode to AUTOCAL (applied
to relay 6 in Fig.)
• Service switch:
Occurrence of a function control.
In this menu you can trigger external solenoid valves via relay
contacts:
• Zero gas:
: relay 8
The zero gas supply which is triggered with AUTOCAL
84124
• Span gas:
The calibration gas supply (assigned to relay 7 in Fig.)
• Sync:
Synchronization of an AUTOCAL with other devices within a
system (assigned to relay 8 in Fig.; see section System setup
with several analyzers in parallel (Page 84)).
Equipment Manual, 04/2022, A5E37100388-012
ULTRAMAT 23