8.2

Calibration

In this function group you can use one or more calibration gases to calibrate the IR channels of
the ULTRAMAT 23 and to readjust the zero point and sensitivity (calibrate the full-scale value).
In addition, you can calibrate the oxygen sensor and the internal atmospheric pressure sensor
and also define the AUTOCAL parameters. The figure shown below indicates the menu sequence
for an analyzer without UV module and H
described in the following sections.
The calibration functions can only be addressed if you have enabled code level 1.
Level 1 required
Calibr. IR channels
Calibr. O2 sensor
Cal. pressure sensor
AUTOCAL/drift values
Thermo AUTOCAL: OFF
Drift values
Cycle time AUTOCAL
Purge time
Purge time
AUTOCAL
Cycle time: XX hours
time left:
Select component
: NNNNN 1
Drift
Zero:
MR1:
Reset drift values
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012
A5E40715586
X.X
8142
82
827
XXXX seconds
8273
XX.XX h
8272
8271
NNNNN
vpm
X
X.X MR2:
X.X
82711
If the device has a UV module, the information about this firm‐
ware can be called up in the additional menu screen (8142).
S sensor. The menu displays and their functions are
2
Select component
O2 cal after install
Start O2 zero cal.
Start O2 span cal.
822
AUTOCAL O2 sens: YES
AUTOCAL with N2: NO
Zero calibr. O2
Span calibr. O2
Span calibration
Setpoint
MR 1+2
Start cal.
MR 1+2
8232
Zero calibration
XX.XX %
O2
Nitrogen flow please
Correct with ENTER
8231
Setpoint: XXXX mbar
Reading:
XXXX mbar
Component
NNNNN
Set span gas values
Start cal. MR 1
Start cal. MR 2
Set sp: XXXXX mg/m
Reading: XXXXX mg/m
Connect span gas
823
Total cal.:OFF NNNNN
with AUTOCAL: no
Span 1: XXXXX mg/m
Span 2: XXXXX mg/m
O2 cal after install
Date :DD.MM.YYYY
O2 zero cal. with N2
XX.XX %
Nitrogen flow please
Correct with ENTER
826
O2 span calibr.
Setpoint
Start cal.
Functions
8.2 Calibration
821
3
3
8212
3
3
8211
8221
O2
8222
MR 1+2
MR 1+2
8223
113

Advertisement

Table of Contents
loading

Table of Contents