Analyzer Status: Status: Maintenance Request; Analyzer Status: Status: Autocal Deviation - Siemens ULTRAMAT 23 Series Equipment Manual

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Functions
8.1 Diagnostics
8.1.1.2

Analyzer status: Status: Maintenance request

AUTOCAL drift
beyond tolerance
Press ENTER to clear
Next message with ->
8.1.1.3

Analyzer status: Status: AUTOCAL deviation

IR drift readings
XXX AUTOCAL
Deviation:
Max.allowed:XXX.XX %
Note
This function is component-specific, but only available for the IR measured components (not for
the UV measured components).
104
This dialog indicates the logged maintenance requests. A main‐
tenance request is set if the values of certain parameters have
reached defined limits, but the analyzer is still able to measure
(e.g. AUTOCAL deviation or O
tion Analyzer status: Status: O2 sensor status (Page 105)). A
corresponding message is output in alphanumeric text.
Access is protected by code level 1.
Following access to the maintenance request list you can:
• Display all recorded faults in succession using the <→> key
• Delete the currently displayed fault message using the <EN‐
TER> key. If present, the next maintenance request is then
displayed. You should delete all maintenance requests
whose causes have been eliminated.
Note
If the analyzer is in measuring mode, you can recognize the
occurrence of a maintenance request in that an "M" appears at
the right edge.
This dialog indicates the deviation in setpoint between several
AUTOCAL procedures. The parameters have the following mean‐
XXX.XX %
ings:
• The text in the two top lines provides information on the
number of AUTOCAL procedures which have been carried
out since the reference value for AUTOCAL was last set (see
section Configuration: Special functions: AUTOCAL deviation
(only for IR components) (Page 146)).
• Deviation is the measured deviation of the actual value from
the reference value, displayed in % of the set measuring
range (with autoranging, range 1 is assumed). This deviation
must not be above the set maximum value.
• Max. allowed is the maximum permissible value for the de‐
viation. Refer to section Configuration: Special functions:
AUTOCAL deviation (only for IR components) (Page 146) for
setting the maximum value.
sensor status; see also sec‐
2
Equipment Manual, 04/2022, A5E37100388-012
ULTRAMAT 23