Technical Specifications; General Technical Data - Siemens ULTRAMAT 23 Series Equipment Manual

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Description

3.4 Technical specifications

3.4
Technical specifications
3.4.1

General technical data

General information
Measured components
Measuring ranges
Characteristics
Operator panel
Operating position
Vibrations
During operation
During transport
Enclosure
Weight
Degree of protection
Electrical characteristics
EMC interference immunity
(with safety extra-low voltage (SELV) with safe
isolation)
Auxiliary power supply
Power consumption
Electrical inputs and outputs
Analog outputs
Relay outputs
Binary inputs
Serial interface
32
Maximum of 4
2 per component, freely adjustable between smallest and largest measur‐
ing range
Linearized
LCD with LED backlighting and contrast control, 80 characters (4 lines à 20
characters); function keys
Front panel vertical
No vibrations
Maximum vibration 5 m/s
2
Approx. 10 kg (22 lbs.) The weight varies according to the ordered variant.
IP40 in accordance with EN 60529 for the versions 7MB235x (TÜV)
IP20 in accordance with EN 60529 for the versions 7MB233x (standard)
In accordance with standard requirements of NAMUR NE21 or
EN 61326-1
100 V AC, +10%/-15%, 50 Hz,
120 V AC, +10%/-15%, 50 Hz,
200 V AC, +10%/-15%, 50 Hz,
230 V AC, +10%/-15%, 50 Hz,
100 V AC, +10%/-15%, 60 Hz,
120 V AC, +10%/-15%, 60 Hz,
230 V AC, +10%/-15%, 60 Hz
Max. 60 VA depending on ordered version
1 analog current output per component, 0/2/4/NAMUR ... 20 mA, floating,
max. load 750 Ω
8, with changeover contacts, freely selectable, e.g. for fault, loading ca‐
pacity 24 V AC/DC/1 A, floating, non-sparking
3, dimensioned for 24 V, floating
Pump
AUTOCAL
Synchronization
ELAN (RS485), PROFIBUS-PA/DP as option