Proceed as follows to install the new sensor:
1. Tighten the new H
Figure 10-3
Position of the H
2. Plug the cable with the connector onto the sensor.
3. Close the analyzer by screwing tight the cover.
4. Switch the analyzer on and wait for the warming-up phase.
5. Check for leaks in the device.
This is described in section Leaks in the gas paths (Page 80)
6. Calibrate the H
7. Enter the installation date with the menu command "H
The analyzer is then ready for use again.
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012
S sensor by hand in the holder up to the mechanical limit
2
S sensor in the analyzer
2
S sensor as described in section Calibration: H2S sensor (Page 120)
2
Service and maintenance
10.2 Maintenance work
S Installation date".
2