3.4.6
Hydrogen sulfide sensor
The cyclic Autocal must be switched on, see also Application note (Page 163).
General information
Measuring ranges
•
Smallest measuring range
•
Largest measuring range
Service life of the sensor
Shelf life
Operation mode
AUTOCAL
Climatic conditions (restrictions)
Permissible ambient temperature
•
During operation
•
During transportation and storage
Permissible ambient pressure
Influencing variables
Associated gases
Interference gases
Drift
Temperature
Air pressure
Measuring behavior at analog output
Output signal noise
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012
0 ... 5 vpm
0 ... 50 vpm
Approx. 12 months after first use in the ULTRAMAT 23
Maximum of 12 months after date of manufacture
Continuous measurement between 0 and 12.5 vpm
Discontinuous measurement between 12.5 and 50 vpm
Cycle time: 1 hour (see Application note (Page 163))
+5 ... +40 °C (41 ... 104 °F)
-10 ... +55 °C (14 ... 131 °F)
750 ... 1200 hPa absolute
The hydrogen sulfide sensor cannot be used if the associated gas contains
the following components:
•
Compounds containing chlorine
•
Compounds containing fluorine
•
Heavy metals
•
Aerosols
•
Alkaline components (e.g. NH
1360 vpm SO
result in a cross-interference of <20 vpm H
2
180 vpm NO result in a cross-interference of <150 vpm H
no cross-interference of CH
< 1% per month
< 3%/10 K based on full-scale value
< 0.2 % of measured value per 1% pressure change
< 2% of smallest measuring range with a damping constant of 30 s
3.4 Technical specifications
>5 mg/m
)
3
3
, CO
and H
(1000 vpm)
4
2
2
Description
S,
2
S,
2
41