Configuration: Special Functions: Factory Data/Reset/Units - Siemens ULTRAMAT 23 Series Equipment Manual

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Configuration: Special functions: ELAN/PROFIBUS/external interference: Cross-interference
Interf. CO2
Cal. w. interf. OFF
Const. interf.
Const.i: +0.0000e+0
8.4.2.4

Configuration: Special functions: Factory data/reset/units

Configuration: Special functions: Factory data/reset/units
Load factory data
Reset analyzer
Change units
Various switches
Configuration: Special functions: Factory data/reset/units: Load factory data
ATTENTION! Data will
be overwritten!
Press ENTER to cont.
Press ESC to cancel
Factory data loading
please wait
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012
After selecting this parameter, you are initially requested to
vpm
determine the component for which this function applies. Af‐
OFF
terwards, the adjacent menu display appears. You can use this
function to:
84234
• Switch the correction of cross-interference on or off for the
duration of the calibration.
To do this, you must select the 2nd line and switch the
parameter on or off there.
– OFF (factory setting) means that the correction of cross-
interference is switched off during the calibration.
– ON means that the corrections of cross-interference re‐
main active during the calibrations. It is thus possible to
use gas combinations as calibration gases.
• Switch the correction of a constant cross-interference on or
off. To do this, you must select the 3rd line and switch the
parameter on or off there. If the constant cross-interference
is switched ON, the measured value of the selected com‐
ponent is corrected with the entered value by adding.
You can use this menu to select a number of items with which e.g.
inappropriate configurations and analyzer settings can be can‐
celled:
The adjacent menu display appears when you select the function
8424
from the higher-level menu.
In this menu you can reestablish the original parameters present
when the analyzer was delivered.
Note:
All modifications (parameters and configuration) which you have
84241
made since then are deleted.
The adjacent display appears when you select this function. You
define the further sequence by pressing either the <ENTER> or
<ESC> key.
When you select this function, the adjacent display appears for
the duration of the load procedure.
84241a
Functions
8.4 Configuration