Spare parts/accessories
13.7 UV module
Part No.
Designation
1 *)
IR source with temperature sensor
2.1
Spacer
2.2 **)
Optical filter
3 *)
Chopper 2-channel
4
Chopper plate glued 2-channel
5.1
O-Ring for analyzer chamber
5.2
Analyzer chamber
6
Gas filter
7.1,7.2 * **)
Receiver chamber
*) Following replacement of this part, special work is required which can only be carried out by a Siemens certified service center
or by qualified, specially trained personnel, for example temperature compensation, basic electronic adjustment, 3-point
calibration with calibration gas etc.
**) Following replacement of this part, the water vapor cross-sensitivity must be checked.
13.7
UV module
Figure 13-19
214
UV module on mounting plate
Remarks
For IR source, if optical filter available
Depends on selected measuring range and measured
component
with threaded rods and windows
required for analyzer chamber lengths 20 mm and
more
Without o-Ring
Depends on selected measuring range and measured
component
Depends on selected measuring range and measured
component
Depends on the selected measured component
Equipment Manual, 04/2022, A5E37100388-012
ULTRAMAT 23