Functions
8.4 Configuration
Configuration: Special functions: ELAN/PROFIBUS/external interference: PROFIBUS parameters
Address
Ident number
PB Relay
Diagnosis
Firmware
Boot FW
TAG: ULTRAMATDREIUND
150
You can use this function to set the following PROFIBUS param‐
: 126
eters:
:
1
: OFF
• Address
This function can be used to set a PROFIBUS station address
84233
to all numerical values between 0 and 126.
• Ident number
: 2.0.0
This parameter is used to set the configuration response of
: 0.2.0
the device. The values 0, 1 and 3 can be set as valid param‐
eters. They have the following meanings:
– 0:
84233a
Only the 'Profile ID number' is positively acknowledged
– 1:
Only the device-specific 'ID number' is positively acknowl‐
edged.
Note:
In order to work with the provided GSD and DD, the 'ID
number' parameter must have the value 1.
– 3:
Only the 'Profile ID number' for multi-variable devices
(complex analyzers) is positively acknowledged.
• PB relay
You can use this function to enable the 8 relays of the add-on
card for control via PROFIBUS. To allow activation, none of
these relays must already be occupied by a device-internal
function.
Note:
The function 'PB relay' is only possible as of PROFIBUS card
firmware version 2.0.0 (shown as Firmware in the figure).
• Diagnosis
If the 'Diagnosis' parameter is selected, the 'Firmware' display
appears with, for example, the following parameters:
– Firmware
The firmware version is displayed here
– Boot FW
The version of the boot firmware is displayed here
– TAG
The name assigned to this analyzer in the network (or the
first 16 characters).
Equipment Manual, 04/2022, A5E37100388-012