Change range SO2
MR 1.0....400 mg/m
MR 2.0...2000 mg/m
ULTRAMAT 23
Equipment Manual, 04/2022, A5E37100388-012
The full-scale values of the measuring ranges can be set in the
third and fourth lines of this dialog. They must lie within the
factory settings, i.e. if an analyzer is factory-set for a total range
3
3
from 0 to 2000 mg/m
this range. The following definitions also apply:
8312
• The smaller MR must not be greater than the higher MR.
• The following input limits apply to the ranges:
– Lower limit: 0.01 times the smaller MR according to fac‐
tory setting (label)
– Upper limit: 1.1 times the higher MR according to factory
setting (label)
In the example:
• Smallest MR 1: 0 to 4 mg/m
• Highest MR 2: 4 to 2200 mg/m
, modifications are only possible within
3
3
3