Siemens ULTRAMAT 23 Series Equipment Manual page 148

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Functions
8.4 Configuration
The parameterization of two analyzers for correction of cross-interference is demonstrated
using the following example. Analyzer 1 delivers the measured values, analyzer 2 uses these
values for a correction calculation.
Note
Neither analyzer is capable of measurements during the AUTOCAL. Therefore it may be
necessary to evaluate signals for the function control.
Example of correction of cross-interference of CO
ELAN
148
on CO with 6 vpm CO at 25% CO
2
Equipment Manual, 04/2022, A5E37100388-012
via
2
ULTRAMAT 23