Siemens ULTRAMAT 23 Series Equipment Manual page 206

Continuous gas analysis
Hide thumbs Also See for ULTRAMAT 23 Series:
Table of Contents

Advertisement

Spare parts/accessories
13.6 IR analyzer units
Part No.
Designation
4
Chopper plate glued 2-channel
5.1
O-Ring for analyzer chamber
5.2
Analyzer chamber
5.3
Analyzer chamber
7.1,7.2 *
Receiver chamber
**)
8 **)
Optical filter with filter support
*) Following replacement of this part, special work is required which can only be carried out by a Siemens certified service center
or by qualified, specially trained personnel, for example temperature compensation, basic electronic adjustment, 3-point
calibration with calibration gas etc.
**) Following replacement of this part, the water vapor cross-sensitivity must be checked.
206
Remarks
with threaded rods and windows
required for analyzer chamber lengths 20 mm and more
Equipment Manual, 04/2022, A5E37100388-012
ULTRAMAT 23

Advertisement

Table of Contents
loading

Table of Contents