PWM Setting

Background
1. Select PWM
setting
2.PWM setting
appears
Note
62
The PWM setting will set the duty of the PWM Drive
setting. The duty is set with ON and OFF times for the
waveform.
OFF time
ON
time
See page 30 for Drive setting details.
From one of the main screens, press
the
ESC
key so that the menu
system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter.
Go to PWM and press Enter.
Use the arrow keys to set the ON and OFF time for the
duty. Press Enter to confirm the setting.
PWM
ON
OFF
*The ON time setting is set in "time units", not
milliseconds. The amount of time in a time unit depends
on the line frequency settings (see page 61).
Line frequency
60Hz
50Hz
Pressing ESC before pressing ENTER will exit the PWM
setting.
T3MIL Series User Manual
03 ~ 99 time units*
0100 ~ 9999 ms
1 Time Unit
16.6mS
20mS