For further details on the ABS or △% compare modes,
see the description in the Compare section, page 38.
Although the 8 bins have their own upper and lower
limits, they still share a common reference value.
Use the arrow keys to go to the Reference setting and
press Enter.
Use the left and right arrow keys to select a digit. Use the
up and down arrow keys to edit the value of the selected
digit and the unit. Press Enter to confirm the setting.
Move
and edit
Select and
Enter
confirm
000.0001~ 999.9999(mΩ/Ω/kΩ/MΩ)
Range
Use the arrow keys to go to the upper limit of the first
bin and press Enter.
Use the Left and Right arrow keys to select a digit. Use
the Up and Down arrow keys to edit the value of the
selected digit and unit. Press the Enter key to confirm the
setting.
Repeat for the lower setting.
Repeat for the remaining bins.
Upper, lower
reference
Move
and edit
Select and
Enter
confirm
ABS mode: 000.0000~999.9999
Setting range
(mΩ/Ω/kΩ/MΩ)
△% mode: -999.99 ~ +999.99