Temperature Compensation

Background
1. Select the
Temperature
Compensation
mode
If the resistance of a DUT at a particular temperature is
needed, the compensation function can be used. This
function can simulate the resistance of a DUT at a
desired temperature. If the ambient temperature and the
temperature coefficient of the DUT are known, it is
possible to determine the resistance of a DUT at any
temperature.
The Temperature Compensation works on the following
formula:
R
t
R
t
0
α
1
(
t-t
)
t
0
0
Where:
R
= Measured resistance value (Ω)
t
R
= Corrected resistance value (Ω)
t0
T
= Inferred absolute temperature
0
t
= Corrected temperature (ºC)
0
t = Current ambient temperature (ºC)
α
= Temperature coefficient of resistance at the correct
to
temperature.
a
to
Press
to access the Temperature Compensation
TC
function.
The temperature-compensated resistance measurement
will appear on the display.
Temperature
compensation
function indicator
Correct Temperature, Temperature Coefficient settings
1
.
T |
|
t
0
0
Extrapolated resistance
measurement at the desired
("correct") temperature
MEASUREMENT
Ambient
temperature
source
Ambient
temperature
49