Status System - Teledyne T3MIL50 User Manual

D.c. milli-ohm meters
Table of Contents

Advertisement

Status system

The diagram below is a description of the status system.
Questionable Data Status
Register
Event
Voltage Overload
0
Not Used
1
Not Used
2
Not Used
3
Not Used
4
Temp Overload
5
Not Used
6
Not Used
7
Not Used
8
Ohms Overload
9
Not Used
10
Limit Test Fail LO
11
Limit Test Fail HI
12
Not Used
13
Not Used
14
Not Used
15
Standard Event Status
Register
Event
0
OPC (Operation Complete)
Not Used
1
QUE (Query Error)
2
Device Error
3
EXE (Execution Error)
4
CME (Command Error)
5
Not Used
6
PON (Power On)
7
For the following command sets, please refer to the diagram above:
STAT: QUES: EVEN?
STAT: QUES: ENAB
STAT: QUES: ENAB?
*ESR?
*ESE
*ESE?
*STB?
*SRE
*SRE?
138
Enable
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
Enable
0
1
2
3
4
5
6
7
T3MIL Series User Manual
Output
Buffer
:
:
:
Status Byte Register
Summary
Not Used
0
Not Used
1
Not Used
2
QUES(Questionable Data)
3
MAV(Message Available)
4
ESB(Standard Event)
5
MSS(Request Service)
6
Not Used
7
Enable
0
1
2
3
4
5
6
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

T3mil50x

Table of Contents