Trigger Delay

Background
Note
1. Select Trigger
Delay setting
58
The Trigger Delay setting adds a delay when an external
trigger signal is recognized. Normally the external trigger
is recognized when there is no contact bounce in the
signal for a fixed length of time, this time is known as the
bounce monitoring window. This ensures that the
external trigger signal is stable before it is recognized.
The Trigger Delay time starts right after the bounce
monitoring window ends.
Bounce monitoring
window
External
Trigger signal
Measurement
process
The Trigger Delay setting is turned off by default.
Pin 2 of the Handler/Scan/Ext I/O interface is used for
external triggering, See page 70 for pinout details.
From one of the main screens, press
the
ESC
key so that the menu
system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter.
Go to Trigger Delay and press
Enter.
T3MIL Series User Manual
Start measurement
Contact
bounce
Trigger
delay time
Measurement
delay time
period
Measurement
period
Measurement
time
Meas. Setup