GND Terminal
GUARD Terminal
Function Keys
Connect the GND (ground) terminal to
GND
the earth ground.
The GUARD terminal has the same
potential as earth, but cannot be
substituted for it. Connect the GUARD
terminal to the cable shield layer of the
GUARD
test leads to help reduce noise.
The Ohm key activates the resistance
Ohm
measurement function.
The Compare key activates the
Compare
comparator function.
The Binning key activates the binning
Binning
function to grade the DUTs into eight
bins according to the tolerance settings.
T3MIL50X only.
The TC key activates the TC
TC
(temperature compensation) function
which calculates the resistance of a
DUT at a specified temperature given
the resistance of the DUT at the
ambient temperature and the
temperature coefficient of the DUT is
known.
The TCONV (Temperature
TCONV
Conversion) function calculates the
temperature of a DUT given an initial
temperature, initial resistance, measured
resistance and a constant (inferred zero
resistance temperature) for the DUT.
The TEMP key activates the
TEMP
temperature measurement function.