Parameter/
Return parameter
Example
TEMPerature:AMBient:DATa
Description
Syntax
Query Syntax
Parameter
Return parameter
Example
SYSTem:LFRequency
Description
Syntax
Query Syntax
Parameter/
Return parameter
<NR1>
OFF
ON
TEMP:AMB:STAT OFF
Disables the user-set ambient temperature.
Sets or returns the user-set ambient temperature value for
the temperature compensation and the temperature
conversion function.
TEMPerature:AMBient:DATa <NRf>
TEMPerature:AMBient:DATa?
<NRf>
<NR2>
TEMP:AMB:DAT 25.6
Sets the user ambient temperature value to +25.6ºC.
TEMP:AMB:DAT?
>25.6
Returns the set ambient temperature as 25.6ºC.
Sets or returns the frequency setting for the line filter.
SYSTem:LFRequency {AUTO | 50 | 60}
SYSTem:LFRequency?
AUTO
50
60
COMMAND OVERVIEW
0:OFF.
1:ON.
Disables the user-set ambient
temperature.
Enables the user-set ambient
temperature.
-50.0~399.9 (Unit: ºC)
-50.0~399.9 (Unit: ºC)
The frequency setting for the line filter
is automatically detected.
The frequency is 50Hz.
The frequency is 60Hz.
Set
Query
Set
Query
125