MEASUREMENT
The △% compare function displays the deviation of the
measured value from the reference value as a percentage.
{[(Measured Value-Reference)/Reference]%}.
Compare function
Measured value
indicator
Deviation of the
measured value from
the reference value as
a percentage
Pass/Fail judgment
Reference, limits, compare
mode and beep mode
The upper (HI) and lower (LO) limits are set as a
percentage from the reference value. (Identical to the %
compare mode)
A measured value that falls within the upper and lower
limits is considered IN (pass), a value that falls below the
lower limits is considered LO, and a value that falls over
the upper limit is a HI.
Lower
Reference
Upper
limit
value
limit
%
LO
IN
HI
The % compare mode displays the measured value as a
percentage of the reference value [(Measured
Value/Reference Value)%].
The upper (HI) and lower (LO) limits are set as a
percentage from the reference value. (Identical to the △
% compare mode)