Teledyne T3MIL50 User Manual page 50

D.c. milli-ohm meters
Table of Contents

Advertisement

2. Ambient
Temperature
3. Temperature
compensation
50
The ambient temperature can be either measured with
the PT-100 sensor or be set manually.
If using the PT-100 sensor the Ambient temperature
setting should be turned off. If the PT-100 probe is not
used, then the ambient temperature needs to be manually
set.
From the bottom menu, go to Meas. Setup > Ambient
Temperature and set the ambient temperature.
See page 60 for setting details.
Range
Off, -50.0 ºC ~ 399.9ºC
Use arrow keys to go to Correct Temperature or to
Temperature Coefficient and press Enter to select the
setting.
To edit the setting values use the left and right arrow keys
to select a digit and use the up and down arrow keys to
edit the digit. Press Enter to confirm the setting.
Move
and edit
Select and
Enter
confirm
Desired Temperature range
Temperature Coefficient range
Below are the inferred zero resistance temperatures of
some common conductors:
Material
Silver
Copper
Gold
Aluminum
Tungsten
Nickel
Iron
T3MIL Series User Manual
Correct temperature, temperature
coefficient settings
-50.0
-9999 ~ +9999 ppm
Inferred Absolute Temperatures
-243
-234.5
-274
-236
-204
-147
-162
~
˚C
+399.9