Teledyne T3MIL50 User Manual page 41

D.c. milli-ohm meters
Table of Contents

Advertisement

3. Reference value
setting
Note
4. Upper & lower
limit setting
Note
Use the arrow keys to navigate to the Reference setting
and press Enter.
Use the left and right arrow keys to select a digit. Use the
up and down arrow keys to edit the value of the selected
digit and the unit. Press Enter to confirm the setting.
Move
and edit
Select and
Enter
confirm
000.0001~ 999.9999 (mΩ/Ω/kΩ/MΩ)
Range:
After setting the Reference value, the displayed
△% values
will be changed to reflect the new Reference
value setting.
Use the arrow keys to navigate to the Upper or Lower
limit setting and press Enter.
Use the left and right arrow keys to select a digit. Use the
up and down arrow keys to edit the value of the selected
digit. Press Enter to confirm the setting.
Repeat for the other limit (Upper or Lower).
Move
and edit
Select and
Enter
confirm
ABS mode: 000.0000~999.9999
Setting Range:
(mΩ/Ω/kΩ/MΩ)
△% and % mode:
-999.99 ~ +999.99
The upper limit must be higher than the lower limit. Not
setting the upper limit higher than the lower limit is not
allowed. Likewise the lower limit cannot be set higher
than the upper limit.
Reference
Upper, Lower reference
MEASUREMENT
△, % or
41