Powerbox; Generator; Errors; Current Errors - Teledyne CP Series User Manual

Table of Contents

Advertisement

User Manual
CP Series
Version 3.1 – June 2020
5.6.1 P
B
OWER
The 'View history' menu will list all exposures recorded by the

PowerBox.

Further detail of the exposure can be viewed by selecting the desired
exposure with the navigation keys (15) and pressing the 'OK' button.
The contents of the list on the PowerBox will differ from that of the
generator which is described in the next chapter since the generator will only record an entry if the exposure
has started. The PowerBox will add a record as soon as the START button is pressed. Exposures that have
been stopped in the prewarning time will thus be recorded by the PowerBox but not by the generator.
The PowerBox can be used in conjunction with different generators so records of exposures with generator
with a different serial number may be listed in the PowerBox but will not be found in the listing of the

generator.

The 'Export to usb' will export the PowerBox's listing to USB when a USB pen drive is plugged in. This
way the entire exposure listing can be extracted from the PowerBox for archiving or failure analyses
purposes.
5.6.2 G
ENERATOR
The 'Show all' menu item will retrieve the exposure listing from the
connected generator and list them on the display.
Further details of a particular exposure can be viewed by navigation to
the desired exposure using the navigation keys (15) and pressing the
'OK' button.
If no generator is connected, no listing can be retrieved, and a padlock symbol will be shown before the
generator menu item.
5.7
E
RRORS
The error messages on the PowerBox can be split up in warnings and
error messages. Warnings will inform the user but are not critical for the
operation of the CP SERIES generator or PowerBox. In total there are
24 warning and error bytes for different categories.
5.7.1 C
URRENT ERRORS
'Current errors' will provide a full text explanation of the current error if the PowerBox or CP SERIES
generator is facing a warning or error situation. Suggestions on how to resolve the problem are displayed
on the main display (1).
5.7.2 E
RROR STATUS
'Error status' will display the current status of all 27 warning and error bytes from all 6 categories in the
form of hexadecimal codes. The warning and error categories are:
OX
PowerBox
View history
Export to USB
Generator
Show history
Export to USB

Errors

Current Errors

Error status

View history
Export to USB
48

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents