Repair Sensor Housing - Emerson Rosemount OCX8800A Manual

Oxygen and combustibles transmitter
Table of Contents

Advertisement

Manual
00809-0500-4880
Install remote electronics housing
Procedure
1. Mount remote electronics housing on wall or pipe within distance of signal
2. Remove the electronics housing cover.
3. If removed, install the power and signal cables and the customer power and
4. Connect the signal cable leads to the O
5. Connect the heater cable leads to the heater power connector (J3).
6. Connect the ground lead to the ground stud. Secure the connection with
7. If used, connect external relay leads to the alarm output relay terminal.
8. Connect the test gas and instrument air lines to the remote electronics
9. Refer to
10. Install the cover on the electronics housing.
11. Turn on the test gases at the cylinders and open the instrument air valve.
12. Restore power to the system.
10.3 

Repair sensor housing

Use the following procedures to remove damaged components from the
transmitter's sensor housing and to install new replacement parts.
Disassemble the transmitter only as needed to replace damaged components.
Use the applicable assembly procedures to install replacement parts and
reassemble the transmitter.
10.3.1 
Disassemble sensor housing
Remove cover and terminals insulator
.
Rosemount OCX8800A
and heater cables in use.
signal conduits and wiring at the electronics housing.
(J4) and to the COe and CJC connector (J5),
two nuts. Attach a separate ground lead (G wire) from the ground stud to
the G terminal on the power input terminal block.
housing. Connect the calibration gas line and instrument air line to the
remote electronics housing.
Figure 10-3
and
electrical connections are complete.
cell and thermocouple connector
2
Figure
Figure
10-6. Make sure all test gas lines and
Maintenance and service
December 2022
10-6.
249