Maintenance and service
December 2022
4. Invert the sensor holder (E,
5. Install and tighten the plug.
268
Figure 10-17: COe Sensor Assembly, Top View
A. Sensor holder flat
B. Sensor holder
C. COe sensor flat
D. Pre-heater flat
E. Sensor holder flat
balls (C) into plug port. Press down on stainless steel balls and tap sensor
holder with plastic hammer to compact balls in pre-heater chamber.
CAUTION
Equipment damage
The resistance temperature device (RTD) elements are fragile, and correct
alignment in sensor holder is required for proper Rosemount OCX8800
operation.
Use care when installing the combustibles (COe) sensor.
Figure